Decay correction is a method of estimating the amount of radioactive decay at some set time before it was actually measured.
Samples may be collected and counted at short time intervals (ex: 1 and 4 hours). But they might be tested for radioactivity all at once. Decay correction is one way of working out what the radioactivity would have been at the time it was taken, rather than at the time it was tested.
For example, the isotope copper-64, commonly used in medical research, has a half-life of 12.7 hours. If you inject a large group of animals at "time zero", but measure the radioactivity in their organs at two later times, the later groups must be "decay corrected" to adjust for the decay that has occurred between the two time points.
where is the original activity count at time zero, is the activity at time "t", "λ" is the decay constant, and "t" is the elapsed time.
The decay constant is where "" is the half-life of the radioactive material of interest.
In this case, "time zero" is Monday, 4:00 p.m., when the first three groups (1,2, and 4 hour animals organs) were measured. The "24 hour" group was measured at 11:00 a.m. Tuesday, which is 19 hours after the first groups.
Start by calculating the decay constant "K". Substitute 12.7 (hours, the half-life of copper-64) for , giving = 0.0546.
Next, multiply this value of "K" by the time elapsed between the first and second measures of radioactivity, 19 hours: (0.0546 x 19) = 1.0368.
Change the sign, to make it -1.0368, then find the "inverse Ln"; in this case 0.3546.
This value is in the denominator of the decay correcting fraction, so it is the same as multiplying the numerator by its inverse (), which is 2.82.
(A simple way to check if you are using the decay correct formula right is to put in the value of the half-life in place of "t". After you perform the inverse Ln, the value should be very close to 0.5. When divided into the value "A" (for uncorrected counts), it effectively doubles them, which is the necessary correction after one half-life has occurred.)
In this case, the uncorrected values will be multiplied by 2.82, which corrects for 19 hours elapsing (between one and two half-lives).
If the radiation measured has dropped by half between the 4 hour sample and the 24 hour sample we might think that the concentration of compound in that organ has dropped by half; but applying the decay correct we see that the concentration is 0.5*2.82 so it has actually increased by 40% in that period.
|
|